CLINITEK Advantus Urine Chemistry AnalyzerGreater Flexibility, Proven Reliability

Respond to demands for higher productivity and high quality with the CLINITEK Advantus® Analyzer. Streamline your workflow with flexible operation. Immediate start-up. Automatic calibration. Network ready. A wide range of options. Added QC features. Flexible operation to meet your needs.

  • Auto-checks for strip identification and humidity exposure improves data integrity, workflow and helps ensure quality results
  • Consistent, reliable results, with one-touch switching between test strip configurations
  • Automatic prompts and test lock-out if QC is not run for improved quality control
  • Automatically read, interpreted, and documented Multistix® test strip results
  • Throughput of 7 seconds/sample; up to 500 tests/hour

Features & Benefits

Reduced risk

  • Avoid errors resulting from reagent degradation with the use of Auto-Checks technology for detection of strip exposure to humidity.*
  • Enhance workflow with Auto-Checks technology for Siemens Healthineers strip type identification.
  • Increase confidence with an advanced quality-control system.

Confidence in results

  • Obtain exceptional result correlation across Siemens Healthineers POC and central-lab urinalysis analyzers.

Fewer errors for reduced follow-up testing

  • Get lab-accurate results for reduced false positives/negatives.
  • Optimize workflow by flagging samples for follow-up microscopy testing.

Broad menu with a small footprint

Flexible productivity for busy clinics and labs

  • Ensure clinical decisions can be made sooner with high throughput.
  • Minimize errors and improve audit trail with convenient bar-code data entry.
  • Improve result interpretation and avoid errors with Auto-Checks technology sample-interference notes.

Network-ready

  • Interfaces with TELCOR, POCcelerator,® UniPOC,™ and RELAYMED Data Management Systems.

Assays

Test Parameters

SG

pH

Protein

Protein Low

Glucose

Ketone

Leukocyte

Bilirubin

Uro-
Bilinogen

Nitrite

Blood

Creatinine

Other

Hema-Combistix®

X

X

X

X

Multistix®

X

X

X

X

X

X

Multistix® 10 SG

X

X

X

X

X

X

X

X

X

X

Multistix® 8 SG

X

X

X

X

X

X

X

X

Multistix PRO® 10 LS

X

X

X

X

X

X

X

 

 

X

X

X

P:C ratio

Technical Specifications

Overview

System Description
Semi-automated urine chemistry analyzer
Tests Measured
Leukocyte, Nitrite, Protein, Blood, Glucose, Ketone, Bilirubin, Urobilinogen, pH, Specific Gravity, Creatinine,‡ and Protein-to-Creatinine Ratio‡
Automatic Measurement
Urine color
Test Format
Dry chemistry reagent strips
Test Measurement
Color change measured by reflectance photometry
Dual readings at reactive and reference wavelengths
Automatically color corrects for Bilirubin
Sample Clarity

Results entered via keyboard or bar code reader

Auto-Checks™ Technology

  • Identify strip types, simplifying training and operation
  • Detect humidity exposure, ensuring data quality
  • Check for common sample-interferences, ensuring result accuracy (not available in the U.S.)

General

Dimensions
12.75 (h) x 15.75 (w) x 13.75 (d) in.
32 (h) x 39 (w) x 35 (d) cm
Weight
7.2 kg (16 lbs)
Power Requirements
100 V to 240 V; 50/60 Hz
Line Leakage Current
<0.5 mA in normal condition
<3.5 mA in single fault condition
Maximum Power Input
72 VA
Ambient Operating Temperature Range
18oC to 30oC (64oF to 86oF)
Ambient Operating Humidity Range
20% to 80% relative humidity
Compliance
UL, CE, EMC

Test Handling

Sample Loading
Self-paced, not dictated by analyzer
STAT Sample Handling
One-touch STAT interrupt capability
Sample Processing
Single or batch mode
Sample ID Entry
Manual, bar code, or downloaded from LIS

Calibration

Calibration
Automatic with every strip

Onboard Computer

Memory
500 patient records plus 200 controls
Interfaces
ASTM (CCS) format and emulates CLINITEK 200 and CLINITEK 200+
Display
Color LCD touchscreen, 5.7 in (14.5 cm), 320 horizontal
by 240 vertical pixels (QVGA)

Computer/Peripheral Interfaces

Ethernet Connection
10/100 Mbit, bidirectional
Onboard Printer
24-character thermal printer
Bar Code
Meets ASTM E1466-92 standards, available with PC
wedge/PS2 connection
Keyboard
External PC keyboard option/PS2 connection
External Printer
Printer interface: parallel port connector (25 pin)
*
Subscription Button Icon
Be the first to know about our events, training, and news